Leírás és Paraméterek
L-3520 Diode Array Detector | |
Diode Array | 1024 pixels |
Wavelength | 190 ~ 640 nm |
Light Source | Deuterium lamp |
Spectral Resolution | 0.6 nm/pixel |
Spectral Width | 4.8 nm |
Wavelength Accuracy | ± 1 nm |
Wavelength Precision | 0.1 nm |
Noise | ± 1.0×10-5 AU (JJG) |
± 0.6×10-5 AU (ASTM) | |
Draft | 0.5×10-3 AU/Hr |
Linear Range | >2.0AU (ASTM) |
Max. Sampling Rate | 8 channels,100 Hz; Full spectrum,100Hz |
Flowcell Pressure | 1200 PSI |
Flowcell Light Path | 10 mm |
Flowcell Volume | 12 μL |
Detection Limit | 2×10-8 g/mL (naphthalene) |
Wavelength Calibration | Mercury peaks and built-in holmium oxide filter |